Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 73
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
PV Bias
-20.00 to 20.00
PPM
-20000 to
20000 PPB
default = 0.000
PV Bias Constant - is used to compensate the input
for drift of an input value.
If PPM Board is installed.
If PPB Board is installed.
Failsafe
0.000 to 20.00
PPM
0.000 to 20000
PPB
default =
20.000
The output value to which the output will go to protect
against the effects of failure of the equipment.
If PPM Board is installed.
If PPB Board is installed.
Filter Time 0 to 120.0
default = 0.0
A software digital filter is provided for dampening the
process. The units are in time constant seconds.