Temperature Input Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 187
Step Action Screen
Use the keys to highlight
the desired “Temperature Input”
selection.
3
Press
Enter
T1 pH/ORP CAL
Temp Cal
Temp Offset
(Read only)
Reset Tmp Offs
T1 pH/ORP CAL
Temp Cal
Temp Offset
(Read only)
Reset Tmp Offs
4
Press
Enter
5 Follow the prompts at the top
and bottom of the screen.
“Place probe in sample”
The display will show the
temperature of the reference
solution as measured by the
probe and Analyzer system.
“Press Enter when stable”
6 Once the reading is stable,
Press
Enter
“Change to sample value”
Use
to change the value to
match the actual temperature of
the reference solution at its
current temperature.
“Enter to save, Exit to cancel”
Limit is ± 5ºC (± 9ºF)
7
Press
Enter
This will save the Temperature
Offset value. If the calibration is
not successful, an error
message will be displayed.
8 If the calibration is suspect,
you can reset the
Temperature Offset and
calibrate again.