SCALE OPERATION
CALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display*
1. Temporarily remove the platter and place it in a safe location.
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and Measures statutes and
regulations applicable in your area by checking with the appropriate government agency before placing a newly calibrated
unit into service or removing any official seals.
2. If this is a currently installed scanner/scale, there will be either an adhesive label seal or wire sealed cover over the scale
calibration button. New installation units will only have a factory-applied adhesive seal covering the scale calibration button
installed.
To remove the adhesive label seal, peel off the label and discard.
To remove the wire seal, cut the wire and discard. Fully loosen the security screw and lift off the cover.
Place the security screw and cover in a safe location.
Figure 42. Diva Scale security seals
3. Press the calibration button to access the Service Mode. ‘Cal then ‘Conf will appear on the remote display.
Figure 43. Enter Service Mode
* These procedures are for MS2320 StratosH Scanner/Diva Scale models only
.