SCALE OPERATION
CALIBRATION
Bar Code Calibration Procedure without Remote Display*
Unit Calibration (Kilograms)
For Pound weight units see instructions on page 43.
Figure 50. Entering Bar Code Calibration Mode (Kilograms)
If a razz tone sounds, an error has occurred. Refer to Diagnostic Indicator Display; Error Codes starting on page 31 for
additional information.
6. Make sure there is no load on the scale platter. The white LEDs will be used to indicate the current step in the calibration process.
7. Calibrate a zero load. The white LEDs will blink once periodically. Wait 8 to 10 seconds for scale stability, and then press the
Multi-Function Button once. After the Multi-Function Button has been pressed, there will be a short delay followed by one short
beep indicating that the zero load value has been stored.
Figure 51.
* These procedures are for MS2320 StratosH Scanner/Diva Scale models only.
Kilograms Calibration